Disclaimer: This is an example of a student written essay.
Click here for sample essays written by our professional writers.

Any scientific information contained within this essay should not be treated as fact, this content is to be used for educational purposes only and may contain factual inaccuracies or be out of date.

Linearity Verification and Evaluation of Within and Between-run Precision of the Albumin Assay Kit

Paper Type: Free Essay Subject: Sciences
Wordcount: 4658 words Published: 23rd Sep 2019

Reference this

Linearity Verification and Evaluation of Within and Between-run Precision of the Albumin Assay Kit 

 

 

Introduction

Albumin is a protein, synthesised in the liver and most abundantly found in plasma. It represents 50% of the body’s total content of protein. It is extremely soluble in water, as a result of it having a negative charge at a physiological pH. It plays a pivotal role in maintaining oncotic pressure. Albumin is involved in the transportation of steroids, calcium ions, ligands and bilirubin. Clinically, high levels of this protein are rarely encountered and are seen in cases of dehydration. Low levels possess greater diagnostic value and can be seen in hepatic disease, gastrointestinal loss, inflammation and ascites. In the Albumin Assay Kit that is utilised in this practical, the albumin binds to the bromocresol green (BCG) indicator. An albumin-BCG complex forms and the absorbance of this is measured spectrophotometrically at 578nm. The concentration of albumin in the sample is directly proportional to the intensity of the colour that is formed by the complex (Caraceni et al., 2013).

Get Help With Your Essay

If you need assistance with writing your essay, our professional essay writing service is here to help!

Essay Writing Service

The term linearity refers to there being a linear relationship between two variables. These variables that are both increasing, are directly proportional and so, the data will be represented in a straight line (Schneider et al., 2010). In the case of this practical, the linearity of the Albumin Assay Kit was evaluated and the manufacturer’s claim of this assay being linear up to 62.2g/l is to be verified. An 80g/l albumin concentration standard was given and dilutions of this standard were made to verify the assay’s linearity. The absorbance of each of the standards were plotted versus the concentration of each standard. The relationship between both variables was then evaluated to test the manufacturer’s claim. The CLSI’s document, EP15-A2 is  intended to verify if the performance of a laboratory is consistent to that claimed by the manufacturer, i.e verification of the precision of the manufacturer’s claim (Chesher, 2008).

Another aspect of this practical that was evaluated is the within-run precision, also known as intra-assay precision and the between-run precision or inter-assay precision, as it is also referred to as. Within-run precision is used to determine the precision of the assay being used. This is carried out by measuring the absorbance of a pooled plasma sample a number of times, assessing the repeatability of the results and so, the precision of the results. The results are assessed on how close they are in agreement to each other after being measured successively under the same conditions. In this practical the sample was measured 10 times. The ideal amount of times a sample should be repeated at a given level, in a single day and in a single run is 20. This will give an idea of the repeatability of the assay. The mean, standard deviation and percentage coefficient of variation were all obtained in order to determine how precise the assay was. Between-run precision was also evaluated. This is a better indicator of the precision of a method than that of within run. This is so as it is a measure of random error of the method. One of these errors being different people carrying out the assay. The EP05-A2, a document produced by the CLSI, is intended to determine the precision of the methods used by the laboratory. This assessment is carried out at two levels, containing at least ten samples per run and should contain at least one sample of quality control (Chesher, 2008). The class results for the albumin concentration (g/l) for both Control 1 and Control 3 were used for this evaluation. Again, the mean, standard deviation and percentage coefficient of variation were all determined and compared to that of the manufacturers. This enabled the comparison of the class results for each control and the manufacturer’s control results in terms of precision

 

Materials

Beaker

Eppendorf tubes

Tube rack

1000µl pipette tips

10µl pipette tips

100 – 1000µl pipette. Manufacturer: Labnet. Serial number: 744060549

2 – 20µl pipette. Manufacturer: Labnet. Serial number: 7044030281

Control 1

Control 3

Calibrator. 32g/l albumin concentration

Saline

Albumin standard

Patient 1 plasma sample

Patient 2 plasma sample

Patient 3 plasma sample

Patient 4 plasma sample

Albumin BCG reagent. Lot number: 3054. Expiry: 2019-04

VWR lab dancer vortex. Model: S040.

Timer. Manufacturer is VWR. Serial number: F153504

Pooled plasma sample

UV spectrophotometer at 600nm

The intensity of the colour that is formed by the complex is directly proportional to the concentration of albumin in the sample.

 

Methods

Table 1: Serial dilution table, showing the concentration of albumin (g/l), the dilution that was carried out for each concentration and the volumes of albumin (µl) and saline, that were used to make up each albumin standard.

Albumin standard (g/l)

Dilution

Volume of albumin (µl)

Volume of saline (µl)

80

8/8

10 x 2 = 20

0

70

7/8 – 7 parts sample + 1 part saline

70 x 2 = 140

10 x 2 = 20

60

6/8 – 6 parts sample  + 2 parts saline

60 x 2 = 120

20 x 2 = 40

40

4/8 / ½ = equal amounts of both sample and saline

50 x 2 = 100

50 x 2 = 100

20

2/8 = 2 part sample + 6 parts saline

20 x 2 = 40

60 x 2 = 120

0

0/8 = 0 parts sample + 8 parts saline

0

10 x 2 = 20

Volumes were multiplied by 2 as 100µl to 200µl of each standard was required in the practical

For all other methods refer to Audit Diagnostics – Albumin BCG Multi-Purpose (MPR) Liquid Reagent kit insert.

 

Results

The following are the results obtained for the albumin colourimetric manual method. The Audit Albumin Kit was used in this practical. The linearity of this kit was assessed to verify the manufacturer’s claim that this assay is in fact linear to 62.2g/l. This kit was also used to evaluate both the within-run precision for the assay, using a pooled plasma sample and the between-run precision of the assay using the class albumin concentrations (g/l) of Control 1 and Control 3. The between-run precision that was obtained was inferior to that of the manufacturer’s, while the within-run precision obtained was quite good.

Tube

Absorbance (A) of first reading (a)

Absorbance (A) of second reading (b)

Average Absorbance (A)

Concentration of albumin (g/l)

Calibrator

0.549

0.552

0.551

32

Control 1

0.545

0.487

0.516

30

80g/l albumin standard

1.406

1.432

1.419

/

70g/l albumin standard

1.279

1.298

1.289

/

60g/l albumin  standard

0.973

1.229

1.101

/

40g/l albumin standard

0.715

0.703

0.709

/

20g/l albumin standard

0.118

0.126

0.122

/

0g/l albumin standard

0.000

/

0.000

/

Patient 1

0.583

0.510

0.547

32

Patient 2

0.447

0.673

0.560

33

Patient 3

0.983

0.981

0.982

57

Control 3

0.931

0.916

0.924

54

Patient 4

0.612

0.604

0.608

35

Table 2: Absorbance (A) of the first reading (a), absorbance (A) of the second reading (b) average absorbance (A)  and albumin concentration (g/l) of the calibrator, control 1, control 3, standards ranging from 0g/l to 80g/l of albumin and the four patient samples using the Audit Albumin Kit colourimetric manual method.  The absorbance readings were taken at 600nm on a spectrophotometer.

Reference range for control 1: 23-33g/l  Reference range for control 3: 39-54g/l.

Both controls are within range and are highlighted yellow to signify this.  

Sample calculation of how to obtain the average absorbance (A) for each sample:

Patient 4 – Absorbance (A) from first reading + Absorbance (A) from second reading / 2                                                                

=    Average absorbance (A)

0.612A+ 0.604A / 2 = 0.608A

Sample calculation of how to obtain the albumin concentrations (g/l) for the controls – C1 and C3 and the patient samples

Albumin concentration = OD of the sample / OD of the calibrator x Concentration of the calibrator.

e.g. C1: 0.516A / 0.551A x 32g/l = 29.97 = 30 g/l.

Figure 1: Linear graph depicting the absorbance (A) of the albumin standards versus their expected concentrations (g/l).

This linear graph verifies the manufacturer’s claim that this albumin assay is indeed linear to 62.2g/l. After this concentration, the linearity ceases and a curve begins to develop

 

 

 

 

 

 

Table 3: Evaluation of the within-run/ intra-assay precision of the Audit Albumin Kit using the manual colourimetric method. Absorbance (A) readings, average absorbance (A) and albumin concentration (g/l) of the calibrator, Control 1, Control 3 and the pooled plasma sample that had its absorbance read ten times as a measurement of within-run precision. A spectrophotometer was used at 600nm. Also included in this table are the mean (g/l), standard deviation and the percentage coefficient of variation (%CV) of the pooled plasma sample.

Tube

Absorbance readings (A)

Average absorbance (A)

Albumin concentration (g/l)

Calibrator

a)0.868           b)0.869

0.869

32

Control 1

a)0.852           b)0.849

0.851

31

Control 3

a)1.264           b)1.329

1.297

48

Pooled plasma 1

1.014

/

37

Pooled plasma 2

1.014

/

37

Pooled plasma 3

1.029

/

38

Pooled plasma 4

1.060

/

39

Pooled plasma 5

1.001

/

37

Pooled plasma 6

1.115

/

41

Pooled plasma 7

1.011

/

37

Pooled plasma 8

1.079

/

40

Pooled plasma 9

1.111

/

41

Pooled plasma 10

1.111

/

41

Mean of pooled plasma replicates (g/l)

38.8

Standard deviation of pooled plasma replicates

1.81

%CV of pooled plasma replicates

4.66

Reference range for control 1: 23-33g/l Reference range for control 3: 39-54g/l.

Both controls are within range and are highlighted yellow to signify this.

%CV= Percentage coefficient of variation

Mean and standard deviation of the pooled plasma replicates was carried out on a Casio fx-83GT Plus calculation in the STAT mode.

Calculation on how to obtain the %CV:      

Standard deviation / Mean (g/l) x100 = %CV.

 

Table 4: Evaluation of the between-run/ inter-assay precision of the Audit Albumin Kit using the manual colourimetric. The class results (1-14 readings) for the albumin concentration (g/l) for Control 1 and Control 3 were used. The reference range (g/l) for each control is included in the table, along with the manufacturer’s and class’s mean (g/l), standard deviation and %CV of Control 1 and Control 3.  The manufacturers took 20 readings of each control.

 

Control 1 albumin concentration (g/l)

Control 3 albumin concentration (g/l)

Reference range (g/l)

23-33

39-54

Manufacturer’s mean  (g/l)

45.35

31.04

Manufacturer’s standard deviation

0.796

0.620

Manufacturer’s (%CV)

1.75

2.00

1

33

52

2

31

49

3

33

52

4

25

  37

5

29

53

6

25

47

7

31

51

8

31

48

9

33

53

10

32

54

11

31

54

12

29

50

13

33

60

14

30

 50 

Class mean (g/l)

30.43

50.71

Class standard deviation

2.68

5.07

Class (%CV)

8.81

10.00

 Two of the class results for Control 3 were outside of the reference range (37 is below and 60 is above the reference range) and are highlighted red to emphasize this.

Discussion

The controls were all within range in the first part of this practical, where the manufacturer’s claim was verified. This can be seen is table 2. As these controls were within range, the results for the assay were valid and so, the assay did not have to be repeated. Again, for the within-run precision assay, as seen in table 3, the controls were all within the reference range. However, for the between-run assay, as shown in table 4, two of the class results for Control 3 were out of range. This could have been due to inaccurate pipetting of the 10µl of the controls or contamination of the control or reagent, causing the absorbance readings to be out of range.

Find Out How UKEssays.com Can Help You!

Our academic experts are ready and waiting to assist with any writing project you may have. From simple essay plans, through to full dissertations, you can guarantee we have a service perfectly matched to your needs.

View our services

From observing figure 1, it is evident that the Albumin Assay Kit is in fact linear up to 62.2g/l, verifying the manufacturer’s claim. After this concentration, linearity appears to dissipate and a curve begins to develop. For this reason, all concentrations higher than 62.2g/l should be diluted, in order to continue to achieve linearity. From observing the graph, it is evident that the 20g/l albumin standard is not in linearity with and does not correlate with the rest of the standards. Reasons for this may be that the exact 10µl volume of the albumin standard was not pipetted accurately to the Eppendorf tube or that the tube was not vortexed prior to incubation, resulting in an erroneous absorbance reading. Contamination could also be the reason for this low albumin concentration. The four patient albumin concentrations fell within the linear range of the albumin standards. However, no further comment can be made on these results as no patient details were given.

In table 3, it is clear that the within-run precision was quite good. A %CV of less than 5 was obtained, emphasising good agreement within the replicates (Westgard.com, 2018). The precision could be further increased and a lower %CV obtained if the pooled plasma sample was repeated until the sample was read 20 times. Hence, this amount of repeatability is a good estimate of within-run precision.

The between-run precision obtained by the class was far below that of the manufacturer. Both controls achieved a %CV of greater than 5, highlighting imprecision in the assay (Westgard.com, 2018).  This is depicted in table 4. However, the manufacturer’s between-run precision is superior as they tested the sample 20 times, as opposed to the class’s 14.  Precision increases upon repeatability of a sample.  As the results that were obtained were that of the class, this could be a reason for the poor precision. Had one person obtained and reported results for all 14 readings of each control, more precision may have been seen.

 

 

 

 

 

 

 

 

Conclusion

In all, the manufacturer’s claim was verified as the Albumin Assay Kit utilised in this practical was linear up to 62.2g/l. Both the within-run and between-run precision were evaluated. The between-run precision obtained by the manufacturer was shown to be far more precise than those obtained by the class, while the within-run precision was shown to be quite precise. The aims have been met and understood for this practical.

 

References

 

Cite This Work

To export a reference to this article please select a referencing stye below:

Reference Copied to Clipboard.
Reference Copied to Clipboard.
Reference Copied to Clipboard.
Reference Copied to Clipboard.
Reference Copied to Clipboard.
Reference Copied to Clipboard.
Reference Copied to Clipboard.

Related Services

View all

DMCA / Removal Request

If you are the original writer of this essay and no longer wish to have your work published on UKEssays.com then please: